Pediatric neuroendoscopy presents specific technical and clinical challenges due to the reduced anatomical space, the narrow operative corridors, and the need to manipulate the endoscope with high precision while preserving a safe pivot at the burr hole. In this context, robotic assistance can improve instrument stability, reduce surgeon fatigue, and enhance motion control, provided that the system remains compact, safe, and compatible with the operating-room workflow. Although robotic assistance has demonstrated value in neurosurgery, especially in adult patients, current systems are often too cumbersome, heavy, and not specifically optimized for direct intraoperative neuroendoscope manipulation in infants and newborns. Therefore, this thesis investigates the design and validation of novel robotic architectures for pediatric neuroendoscopy. A key requirement in this context is the preservation of a Remote Center of Motion, RCM, at the burr hole. Maintaining a fixed and accurate pivot at the cranial entry point facilitates alignment between the endoscope insertion trajectory and the trajectory selected during pre-operative planning. Consequently, the RCM mechanism should also provide sufficient rigidity during insertion, while allowing very small and safe orientation adjustments once the endoscope is already inside the brain. Based on these needs, this thesis places particular focus on compact RCM mechanisms that can satisfy the requirements of miniaturization, sterility, neuronavigation compatibility, safe operation in constrained spaces, and predictable mechanical behavior. The first part of the thesis presents PedRo, Pediatric Robot for Neuroendoscopy, a rigid-body robotic platform specifically designed for patients aged 0 to 3 years. The main contributions of this part lie in the quantitative derivation of pediatric neuroendoscopic requirements, the definition of a compact three-DoF RCM architecture, and the experimental assessment of joint-level positioning and RCM preservation. The robot integrates independently actuated pitch, yaw, and insertion DoFs through curvilinear guides and screw-driven actuation. The design provides rotational ranges of ±20° in pitch and yaw, a 135 mm insertion stroke, a total footprint of 390 mm × 215 mm × 190 mm, and a mass of 3.4 kg. These values exceed the identified clinical requirements. PedRo was experimentally validated through ISO 230-2 point-to-point positioning tests performed using an Aurora electromagnetic tracking system. The results showed precise positioning behavior of all three joints, achieving sub-millimetric insertion resolution and limited RCM displacement well below the 1 mm clinical threshold. The second part of the thesis explores compliant mechanisms as an alternative to conventional rigid-link architectures, with the aim of reducing footprint, backlash, friction, and mechanical complexity. First, the TETRA II joint was redesigned to improve surgical usability by relocating the end-effector outside the structure, thus facilitating sterile draping, quick release, and visibility of the entry point. Subsequently, a dual-joint compliant architecture was developed to overcome stiffness anisotropy, achieving near-isotropic behavior and very low parasitic RCM displacement. This part of the work combined hybrid analytical/FEM modeling, angular-offset optimization, and experimental characterization of a PA12 prototype under controlled radial loads. The optimized design of this mechanism successfully achieved near-isotropic stiffness. Experimentally, a fabricated prototype showed a root-mean-square error of 0.90 N/m with respect to the ideal isotropic case, stiffness values within 19% of FEM predictions, and a maximum parasitic RCM displacement of 0.051 mm for a 4.5° shaft rotation, again well below the 1 mm tolerance from the clinical requirements. Finally, the thesis investigates the design of an off-axis monolithic compliant RCM joint combining near-isotropic stiffness, low parasitic error, compactness, and fabrication readiness in a single-stage architecture. The scientific contributions of this part include the formulation of a two-step design method in which the mobility panels are first dimensioned through an analytical isotropy-oriented criterion, and the constraining panels are then synthesized through FEM feasibility analysis to balance stiffness uniformity, RCM fidelity, and workspace. The design was evaluated numerically and experimentally using an SLS-fabricated PA12 prototype and electromagnetic sensing. The selected configuration achieved a stiffness-ellipse principal-axis ratio, PAR, below 1.37, a parasitic-to-useful rotation ratio, PRR, of 0.0063, and sub-millimetric RCM drift below 0.172 mm for 4.5° rotations. In addition, the device presented a fatigue-bounded workspace ranging from 12.1° to 34.4°, which exceeds the clinical target of ±15° in most directions. Overall, this thesis provides a progressive design pathway from a conventional compact robotic platform to advanced compliant RCM mechanisms for pediatric neuroendoscopy. The results support the feasibility of smaller, safer, and more accurate robotic assistants tailored to the specific constraints of neurosurgical procedures in patients aged 0 to 3 years.

La neuroendoscopia pediatrica presenta specifiche sfide tecniche e cliniche dovute allo spazio anatomico ridotto, ai corridoi operativi ristretti e alla necessità di manipolare l’endoscopio con elevata precisione, mantenendo al tempo stesso un pivot sicuro in corrispondenza del foro di accesso cranico. In questo contesto, l’assistenza robotica può migliorare la stabilità dello strumento, ridurre l’affaticamento del chirurgo e aumentare il controllo del movimento, a condizione che il sistema rimanga compatto, sicuro e compatibile con il flusso di lavoro della sala operatoria. Sebbene l’assistenza robotica abbia dimostrato il proprio valore in neurochirurgia, soprattutto nei pazienti adulti, i sistemi attuali sono spesso troppo ingombranti, pesanti e non specificamente ottimizzati per la manipolazione intraoperatoria diretta del neuroendoscopio in lattanti e neonati. Pertanto, questa tesi studia la progettazione e la validazione di nuove architetture robotiche per la neuroendoscopia pediatrica. Un requisito fondamentale in questo contesto è la conservazione di un Remote Center of Motion, RCM, in corrispondenza del foro di accesso cranico. Il mantenimento di un pivot fisso e accurato nel punto di ingresso cranico facilita l’allineamento tra la traiettoria di inserimento dell’endoscopio e la traiettoria selezionata durante la pianificazione preoperatoria. Di conseguenza, il meccanismo RCM dovrebbe anche garantire una rigidità sufficiente durante l’inserimento, consentendo al tempo stesso regolazioni di orientamento molto piccole e sicure una volta che l’endoscopio si trova già all’interno del cervello. Sulla base di queste esigenze, questa tesi si concentra in particolare su meccanismi RCM compatti, in grado di soddisfare i requisiti di miniaturizzazione, sterilità, compatibilità con la neuronavigazione, funzionamento sicuro in spazi vincolati e comportamento meccanico prevedibile. La prima parte della tesi presenta PedRo, Pediatric Robot for Neuroendoscopy, una piattaforma robotica a corpi rigidi progettata specificamente per pazienti di età compresa tra 0 e 3 anni. I principali contributi di questa parte riguardano la derivazione quantitativa dei requisiti neuroendoscopici pediatrici, la definizione di un’architettura RCM compatta a tre gradi di libertà, DoF, e la valutazione sperimentale del posizionamento a livello dei giunti e della conservazione dell’RCM. Il robot integra movimenti di pitch, yaw e inserimento attuati in modo indipendente attraverso guide curvilinee e attuazione a vite. Il progetto fornisce intervalli di rotazione pari a ±20° in pitch e yaw, una corsa di inserimento di 135 mm, un ingombro complessivo di 390 mm × 215 mm × 190 mm e una massa di 3.4 kg. Questi valori superano i requisiti clinici identificati. PedRo è stato validato sperimentalmente mediante test di posizionamento punto-punto secondo la norma ISO 230-2, eseguiti utilizzando un sistema di tracciamento elettromagnetico Aurora. I risultati hanno mostrato un comportamento di posizionamento preciso per tutti e tre i giunti, ottenendo una risoluzione di inserimento submillimetrica e uno spostamento limitato dell’RCM, ben al di sotto della soglia clinica di 1 mm. La seconda parte della tesi esplora i meccanismi compliant come alternativa alle architetture convenzionali a link rigidi, con l’obiettivo di ridurre ingombro, gioco meccanico, attrito e complessità meccanica. In primo luogo, il giunto TETRA II è stato riprogettato per migliorare l’usabilità chirurgica, spostando l’end-effector all’esterno della struttura e facilitando così il drappeggio sterile, il rilascio rapido e la visibilità del punto di ingresso. Successivamente, è stata sviluppata un’architettura compliant a doppio giunto per superare l’anisotropia di rigidezza, ottenendo un comportamento quasi isotropo e uno spostamento parassita dell’RCM molto ridotto. Questa parte del lavoro ha combinato modellazione ibrida analitica/FEM, ottimizzazione dell’offset angolare e caratterizzazione sperimentale di un prototipo in PA12 sottoposto a carichi radiali controllati. Il progetto ottimizzato di questo meccanismo ha raggiunto con successo una rigidezza quasi isotropa. Sperimentalmente, il prototipo fabbricato ha mostrato un errore quadratico medio di 0.90 N/m rispetto al caso isotropo ideale, valori di rigidezza entro il 19% delle previsioni FEM e uno spostamento parassita massimo dell’RCM pari a 0.051 mm per una rotazione dello strumento di 4.5°, ancora una volta ben al di sotto della tolleranza di 1 mm derivante dai requisiti clinici. Infine, la tesi studia la progettazione di un giunto RCM fuori-asse compliant monolitico che combina rigidezza quasi isotropa, basso errore parassita, compattezza e predisposizione alla fabbricazione in un’architettura a singolo stadio. I contributi scientifici di questa parte includono la formulazione di un metodo di progettazione in due fasi, nel quale i pannelli di mobilità vengono prima dimensionati mediante un criterio analitico orientato all’isotropia, mentre i pannelli di vincolo vengono successivamente sintetizzati attraverso un’analisi di fattibilità FEM, al fine di bilanciare uniformità di rigidezza, fedeltà dell’RCM e spazio di lavoro. Il progetto è stato valutato numericamente e sperimentalmente utilizzando un prototipo in PA12 fabbricato mediante SLS e sensori elettromagnetici. La configurazione selezionata ha raggiunto un rapporto tra gli assi principali dell’ellisse di rigidezza, PAR, inferiore a 1.37, un rapporto tra rotazione parassita e rotazione utile, PRR, pari a 0.0063, e una deriva submillimetrica dell’RCM inferiore a 0.172 mm per rotazioni di 4.5°. Inoltre, il dispositivo ha mostrato uno spazio di lavoro limitato a fatica compreso tra 12.1° e 34.4°, superando il target clinico di ±15° nella maggior parte delle direzioni. Nel complesso, questa tesi fornisce un percorso progettuale progressivo che va da una piattaforma robotica convenzionale e compatta fino a meccanismi RCM cedevoli avanzati per la neuroendoscopia pediatrica. I risultati supportano la fattibilità di assistenti robotici più piccoli, più sicuri e più accurati, progettati specificamente per i vincoli delle procedure neurochirurgiche in pazienti di età compresa tra 0 e 3 anni.

Design and development of a novel robot for pediatric neuroendoscopy

MARIANO, FEDERICO
2026

Abstract

Pediatric neuroendoscopy presents specific technical and clinical challenges due to the reduced anatomical space, the narrow operative corridors, and the need to manipulate the endoscope with high precision while preserving a safe pivot at the burr hole. In this context, robotic assistance can improve instrument stability, reduce surgeon fatigue, and enhance motion control, provided that the system remains compact, safe, and compatible with the operating-room workflow. Although robotic assistance has demonstrated value in neurosurgery, especially in adult patients, current systems are often too cumbersome, heavy, and not specifically optimized for direct intraoperative neuroendoscope manipulation in infants and newborns. Therefore, this thesis investigates the design and validation of novel robotic architectures for pediatric neuroendoscopy. A key requirement in this context is the preservation of a Remote Center of Motion, RCM, at the burr hole. Maintaining a fixed and accurate pivot at the cranial entry point facilitates alignment between the endoscope insertion trajectory and the trajectory selected during pre-operative planning. Consequently, the RCM mechanism should also provide sufficient rigidity during insertion, while allowing very small and safe orientation adjustments once the endoscope is already inside the brain. Based on these needs, this thesis places particular focus on compact RCM mechanisms that can satisfy the requirements of miniaturization, sterility, neuronavigation compatibility, safe operation in constrained spaces, and predictable mechanical behavior. The first part of the thesis presents PedRo, Pediatric Robot for Neuroendoscopy, a rigid-body robotic platform specifically designed for patients aged 0 to 3 years. The main contributions of this part lie in the quantitative derivation of pediatric neuroendoscopic requirements, the definition of a compact three-DoF RCM architecture, and the experimental assessment of joint-level positioning and RCM preservation. The robot integrates independently actuated pitch, yaw, and insertion DoFs through curvilinear guides and screw-driven actuation. The design provides rotational ranges of ±20° in pitch and yaw, a 135 mm insertion stroke, a total footprint of 390 mm × 215 mm × 190 mm, and a mass of 3.4 kg. These values exceed the identified clinical requirements. PedRo was experimentally validated through ISO 230-2 point-to-point positioning tests performed using an Aurora electromagnetic tracking system. The results showed precise positioning behavior of all three joints, achieving sub-millimetric insertion resolution and limited RCM displacement well below the 1 mm clinical threshold. The second part of the thesis explores compliant mechanisms as an alternative to conventional rigid-link architectures, with the aim of reducing footprint, backlash, friction, and mechanical complexity. First, the TETRA II joint was redesigned to improve surgical usability by relocating the end-effector outside the structure, thus facilitating sterile draping, quick release, and visibility of the entry point. Subsequently, a dual-joint compliant architecture was developed to overcome stiffness anisotropy, achieving near-isotropic behavior and very low parasitic RCM displacement. This part of the work combined hybrid analytical/FEM modeling, angular-offset optimization, and experimental characterization of a PA12 prototype under controlled radial loads. The optimized design of this mechanism successfully achieved near-isotropic stiffness. Experimentally, a fabricated prototype showed a root-mean-square error of 0.90 N/m with respect to the ideal isotropic case, stiffness values within 19% of FEM predictions, and a maximum parasitic RCM displacement of 0.051 mm for a 4.5° shaft rotation, again well below the 1 mm tolerance from the clinical requirements. Finally, the thesis investigates the design of an off-axis monolithic compliant RCM joint combining near-isotropic stiffness, low parasitic error, compactness, and fabrication readiness in a single-stage architecture. The scientific contributions of this part include the formulation of a two-step design method in which the mobility panels are first dimensioned through an analytical isotropy-oriented criterion, and the constraining panels are then synthesized through FEM feasibility analysis to balance stiffness uniformity, RCM fidelity, and workspace. The design was evaluated numerically and experimentally using an SLS-fabricated PA12 prototype and electromagnetic sensing. The selected configuration achieved a stiffness-ellipse principal-axis ratio, PAR, below 1.37, a parasitic-to-useful rotation ratio, PRR, of 0.0063, and sub-millimetric RCM drift below 0.172 mm for 4.5° rotations. In addition, the device presented a fatigue-bounded workspace ranging from 12.1° to 34.4°, which exceeds the clinical target of ±15° in most directions. Overall, this thesis provides a progressive design pathway from a conventional compact robotic platform to advanced compliant RCM mechanisms for pediatric neuroendoscopy. The results support the feasibility of smaller, safer, and more accurate robotic assistants tailored to the specific constraints of neurosurgical procedures in patients aged 0 to 3 years.
14-lug-2026
Inglese
La neuroendoscopia pediatrica presenta specifiche sfide tecniche e cliniche dovute allo spazio anatomico ridotto, ai corridoi operativi ristretti e alla necessità di manipolare l’endoscopio con elevata precisione, mantenendo al tempo stesso un pivot sicuro in corrispondenza del foro di accesso cranico. In questo contesto, l’assistenza robotica può migliorare la stabilità dello strumento, ridurre l’affaticamento del chirurgo e aumentare il controllo del movimento, a condizione che il sistema rimanga compatto, sicuro e compatibile con il flusso di lavoro della sala operatoria. Sebbene l’assistenza robotica abbia dimostrato il proprio valore in neurochirurgia, soprattutto nei pazienti adulti, i sistemi attuali sono spesso troppo ingombranti, pesanti e non specificamente ottimizzati per la manipolazione intraoperatoria diretta del neuroendoscopio in lattanti e neonati. Pertanto, questa tesi studia la progettazione e la validazione di nuove architetture robotiche per la neuroendoscopia pediatrica. Un requisito fondamentale in questo contesto è la conservazione di un Remote Center of Motion, RCM, in corrispondenza del foro di accesso cranico. Il mantenimento di un pivot fisso e accurato nel punto di ingresso cranico facilita l’allineamento tra la traiettoria di inserimento dell’endoscopio e la traiettoria selezionata durante la pianificazione preoperatoria. Di conseguenza, il meccanismo RCM dovrebbe anche garantire una rigidità sufficiente durante l’inserimento, consentendo al tempo stesso regolazioni di orientamento molto piccole e sicure una volta che l’endoscopio si trova già all’interno del cervello. Sulla base di queste esigenze, questa tesi si concentra in particolare su meccanismi RCM compatti, in grado di soddisfare i requisiti di miniaturizzazione, sterilità, compatibilità con la neuronavigazione, funzionamento sicuro in spazi vincolati e comportamento meccanico prevedibile. La prima parte della tesi presenta PedRo, Pediatric Robot for Neuroendoscopy, una piattaforma robotica a corpi rigidi progettata specificamente per pazienti di età compresa tra 0 e 3 anni. I principali contributi di questa parte riguardano la derivazione quantitativa dei requisiti neuroendoscopici pediatrici, la definizione di un’architettura RCM compatta a tre gradi di libertà, DoF, e la valutazione sperimentale del posizionamento a livello dei giunti e della conservazione dell’RCM. Il robot integra movimenti di pitch, yaw e inserimento attuati in modo indipendente attraverso guide curvilinee e attuazione a vite. Il progetto fornisce intervalli di rotazione pari a ±20° in pitch e yaw, una corsa di inserimento di 135 mm, un ingombro complessivo di 390 mm × 215 mm × 190 mm e una massa di 3.4 kg. Questi valori superano i requisiti clinici identificati. PedRo è stato validato sperimentalmente mediante test di posizionamento punto-punto secondo la norma ISO 230-2, eseguiti utilizzando un sistema di tracciamento elettromagnetico Aurora. I risultati hanno mostrato un comportamento di posizionamento preciso per tutti e tre i giunti, ottenendo una risoluzione di inserimento submillimetrica e uno spostamento limitato dell’RCM, ben al di sotto della soglia clinica di 1 mm. La seconda parte della tesi esplora i meccanismi compliant come alternativa alle architetture convenzionali a link rigidi, con l’obiettivo di ridurre ingombro, gioco meccanico, attrito e complessità meccanica. In primo luogo, il giunto TETRA II è stato riprogettato per migliorare l’usabilità chirurgica, spostando l’end-effector all’esterno della struttura e facilitando così il drappeggio sterile, il rilascio rapido e la visibilità del punto di ingresso. Successivamente, è stata sviluppata un’architettura compliant a doppio giunto per superare l’anisotropia di rigidezza, ottenendo un comportamento quasi isotropo e uno spostamento parassita dell’RCM molto ridotto. Questa parte del lavoro ha combinato modellazione ibrida analitica/FEM, ottimizzazione dell’offset angolare e caratterizzazione sperimentale di un prototipo in PA12 sottoposto a carichi radiali controllati. Il progetto ottimizzato di questo meccanismo ha raggiunto con successo una rigidezza quasi isotropa. Sperimentalmente, il prototipo fabbricato ha mostrato un errore quadratico medio di 0.90 N/m rispetto al caso isotropo ideale, valori di rigidezza entro il 19% delle previsioni FEM e uno spostamento parassita massimo dell’RCM pari a 0.051 mm per una rotazione dello strumento di 4.5°, ancora una volta ben al di sotto della tolleranza di 1 mm derivante dai requisiti clinici. Infine, la tesi studia la progettazione di un giunto RCM fuori-asse compliant monolitico che combina rigidezza quasi isotropa, basso errore parassita, compattezza e predisposizione alla fabbricazione in un’architettura a singolo stadio. I contributi scientifici di questa parte includono la formulazione di un metodo di progettazione in due fasi, nel quale i pannelli di mobilità vengono prima dimensionati mediante un criterio analitico orientato all’isotropia, mentre i pannelli di vincolo vengono successivamente sintetizzati attraverso un’analisi di fattibilità FEM, al fine di bilanciare uniformità di rigidezza, fedeltà dell’RCM e spazio di lavoro. Il progetto è stato valutato numericamente e sperimentalmente utilizzando un prototipo in PA12 fabbricato mediante SLS e sensori elettromagnetici. La configurazione selezionata ha raggiunto un rapporto tra gli assi principali dell’ellisse di rigidezza, PAR, inferiore a 1.37, un rapporto tra rotazione parassita e rotazione utile, PRR, pari a 0.0063, e una deriva submillimetrica dell’RCM inferiore a 0.172 mm per rotazioni di 4.5°. Inoltre, il dispositivo ha mostrato uno spazio di lavoro limitato a fatica compreso tra 12.1° e 34.4°, superando il target clinico di ±15° nella maggior parte delle direzioni. Nel complesso, questa tesi fornisce un percorso progettuale progressivo che va da una piattaforma robotica convenzionale e compatta fino a meccanismi RCM cedevoli avanzati per la neuroendoscopia pediatrica. I risultati supportano la fattibilità di assistenti robotici più piccoli, più sicuri e più accurati, progettati specificamente per i vincoli delle procedure neurochirurgiche in pazienti di età compresa tra 0 e 3 anni.
Mattos, Leonardo S.
Berselli, Giovanni
Marzullo, Aldo
File in questo prodotto:
File Dimensione Formato  
Mariano Federico - thesis.pdf

accesso aperto

Licenza: Tutti i diritti riservati
Dimensione 72.81 MB
Formato Adobe PDF
72.81 MB Adobe PDF Visualizza/Apri

I documenti in UNITESI sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.14242/384787
Il codice NBN di questa tesi è URN:NBN:IT:POLIMI-384787